EasyManua.ls Logo

Sony Trinitron PVM-2950QM - Line Crawling Adjustment; SECAM Bell Filter Adjustment

Sony Trinitron PVM-2950QM
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PVM-2950Q/2950QM
RM-854
7-6. Connect the oscilloscope to C417 and R372 and adjust
RY314 so that the waveform is as in the figure below.
-------f
0.26 ± 0.02V
---------------__ ...,_____ __ j
Fig. 6-12
7-7. Change the signal to PAL 75% color bars.
7-8. Connect the oscilloscope to CN305 Pin @ and adjust
RY312 so that the waveform is as in the figure below.
__________________________ l
50 mVp-p max.
[ --
Fig. 6-13
8. Line crawling adjustment
8-1. Input 75% PAL color bars from Line 1.
8-2. Connect the oscilloscope to CN305 Pin @ and check that
the output difference per lH for the waveform is under 5%.
8-3. If the difference is over 5%, measure between C416 and
R373 and between C417 and R372, change the signal to a
PAL SP CB signal and adjust Tl on CP301 to minimize the
level difference per lH of the anti-PAL signal.
Anti-PAL signal
t
Fig. 6-14
8-4. Repeat the adjustment from 7-1.
9. SECAM bell filter adjustment
9-1. Input SECAM color bars to Line 1.
9-2. Connect the oscilloscope to IC303 Pin ® and adjust T301
so that the waveform is as in the figure below.
Fig. 6-15
9-3. Input SECAM color bars to Line 1 (100% white).
9-4. Connect the oscilloscope to the emitter of Q359 and adjust
with RY313 so that the waveform is as in the figure below.
0.95 ± 0.05Vp-p.
Fig. 6-16
9-5. Connect the oscilloscope between C417 and R372 and
adjust LY301 so that the B-Y waveform no-color
component level is a straight line.
,..,
I"-,
' '
' '
' '
Fig. 6-17
- 34-

Related product manuals