EasyManua.ls Logo

Spiricon LBA-300 - LOG - Logging

Default Icon
281 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Centroid X,
Centroid Y,
Width X,
Width Y,
Diameter(^END)
LBA-PC sends
RDR
,%,,,um,um,um,um,um,um,um(^END)
10.6.2.6 LOG - Logging
You can configure the LBA-PC to automatically send a FRM or RDD and RDR response each time
new data is acquired. Remote logging is the fastest method available to transfer new frames
from the LBA-PC.
Use the File | Logging… dialog box or the LOG configuration command to specify logging new
data or results to the host. If the data log file name is “FRM” or “RDD” then each time a new
frame of data is acquired the LBA-PC will automatically send a response as if you had sent
“:FRM?” or “:RDD?” respectively. If the results log file name is “RDR” then each time a new
frame of data is acquired the LBA-PC will automatically send a response as if you had sent
“:RDR?”.
The LBA-PC will not capture another frame until the host reads the current logging response.
This example describes how to enable data file logging to the host controller.
Host sends
:LOG
DataLogging=1;
DataFileName=FRM(^END)
Host sends
:RUN
LBA-PC sends
FRM
FrameNumber=1;
#516384(DAB)…(DAB) (^END)
LBA-PC sends
Operator’s Manual LBA-PC
Doc. No. 10654-001, Rev 4.10
197
FRM
FrameNumber=2;

Table of Contents