Command Meaning Usage
Operator’s Manual LBA-PC
Doc. No. 10654-001, Rev 4.10
213
exists (STB & SRE).
*ESE Event Status
Enable
Sets mask to enable event status
notification of corresponding bits set in
ESR.
*ESE? Return ESE contents.
*ESR? Event Status
Register
Return and clear ESR contents:
bit 7 - Power on.
bit 6 - User Request (unused).
bit 5 - Command Error fla
g
, set when host
sends a command that is not recognized.
bit 4 - Execution Error flag, set when the
host commands the LBA-PC to perform a
task which is preempted by other
conditions (e.g. some configurations
cannot be set while running, path does
not exist, illegal file name, etc.)
bit 3 - Device Dependent Error flag
(unused).
bit 2 - Query Error flag, data in output
queue has been lost.
bit 1 - Request Control (unused).
bit 0 - Operation Complete flag, set when:
Ultracal done (only if Ultracal successful)
AutoExposure done
Generate Gain done
Block Capture done
Logging done
Post Process done (only if stop not caused
by error)
Statistics done
Print done
*SRE Service
Request
Enable
Set mask to enable service request
notification of corresponding bits in STB.
*SRE? Return SRE contents.
*STB? Status Byte Returns Status Byte contents:
bit 7 - unused
bit 6 - Master Summary Status, MSS, set
whenever an unmasked status condition