Operator’s Manual LBA-PC
234
Total B,B Min,Max test enable.
A.5.2.6 PSW - enter password
:PSW <configuration>
Key Type Value Description
Password S See online help under Password Lockout for
information about passwords.
:PSW?
Returns:
PSW < configuration >
Key Type Value Description
Lockout B true = local lockout active
A.5.3 Pass/Fail Menu
A.5.3.1 PFF - set pass/fail master configuration
:PFF <configuration>
1
:PFF?
2
Key Type Value Description
PassFail B Master Pass/Fail enable.
TTL B TTL output on ‘When’ condition
enable.
Beep B Beep on ‘When’ condition enable.
Stop B Stop on ‘When’ condition enable.
When L Pass/Fail condition that invokes TTL,
Beep, Stop.
0 = fail
1 = pass
A.5.3.2 pass/fail quant configuration
Key Type Value Description