1) Verify that LRV and URV values are the ones you
want;
2) In the start-scale process condition, use OFFSET
command to correct the measure to zero;
3) In the full scale process condition, use SET S. URV
(set span URV) command to correct the URV value;
It is possible to correct only the analogical ouput, by
changing the LRV or URV values (alternatively editing SET
LRV and SET URV) to set the instrument measure at 4 mA
or 20 mA.
Uncalibrated
analog output
Use the LOOP TEST command to verify the analog output
calibration. If it is not correct, use the CAL4-20mA
command to calibrate the output.
Use the ADDRESS command to verify that the polling
address is set to 0.
Analog output
<4mA o >20mA
Transmitter in
fault mode
Check if the output isn’t out of the allowed span or check
the presence of a failure (e1, e2, … on display).
Check the settings using the ALARM TYPE command.
Set an input average using the DAMP VALUE command (i.e.
10 seconds).
Impossible to
access the
menu
Wrong /
forgotten
password