EasyManua.ls Logo

STATUS SCIENTIFIC CONTROLS MCU3 - 6 Trouble Shooting; Common Faults

Default Icon
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TD06/001
Issue:
6
Change Note:
1696
Page 6-1
6 TROUBLE SHOOTING
The MCU system is supplied fully tested, and it is unlikely that any apparent faults are as a result
of faulty hardware. Most problems encountered are as a result of installation faults or incorrect
software/hardware configuration.
6.1 Common Faults
Fault
The Detector head display is blank.
Ensure the MCU has power connected.
Ensure the current loop connections are correct at the detector head and at the MCU.
Ensure all connectors within the detector head are attached.
Ensure a voltage is present at the loop current connections at the detector head terminal
block.
Unless the detector head is mounted a significant distance from the MCU this voltage
should be approaching 20V (7V minimum).
Where an I.S. Barrier is fitted, check that the barrier fuse is intact.
Fault
A Flammable Head is attached to channel 1 but the reading on the Detector Head display is
a high negative number (e.g. 60).
Check that the pellistor supply is connected to the head.
The flammable heads require a separate supply to provide the pellistors with current in
excess of the 4-20mA available loop current.
Check that the ‘pwr’ supply measures 3.5V 4.5V at the detector head.
The voltage supply for the pellistors is regulated within the detector head at a factory pre-
set level (2.0V for VQ2 type pellistors). In order to provide a regulated supply at the
correct voltage the detector head ‘pwr’ connection must receive a minimum of 3.5V.
Fault
The detector head displays a valid reading; however the MCU does not appear to be
displaying the same, or any valid reading.
Ensure that the channel has been enabled within the software.
Ensure the channel setup is correct i.e. gas type, range
Check that the jumper switch settings are correct. For all Status Scientific Detector Heads J6,
J7 & J11 need to be switched ON.
Calibrate channel.
Fault
The MCU does not appear to communicate with channels five to eight.
Ensure interconnecting cable is correctly attached between upper and lower platforms within
the control unit.
Ensure relevant channels are enabled within the software.

Table of Contents