8.8 FGD Detector Head Calibration ‘Look-up’ Table
Each time a calibration is performed on a detector head, the user will be displayed a zero and a
span factor (refer to section 5.4 for calibration instructions). The values given for these factors
should be compared to the table below to determine the serviceability of the sensor.
Status Scientific Controls strongly advise that sensors providing zero and span factors that do not
correspond to those listed as acceptable must not be used. A new sensor MUST be fitted in the
detector to allow continued accurate and reliable readings to be made (refer to section 7.3 for
sensor replacement instructions).
Sensor
Zero
Span
Min
Typ
Max
Min
Typ
Max
LEL CH
4
50
300
600
200
320
960
O
2
0
5
20
350
450
1300
H
2
S
350
500
600
100
n/d
1500
CO
350
500
600
100
n/d
1500
CL
2
350
500
600
100
n/d
1500
NO
2
350
500
600
100
n/d
1500
n/d : no data.
This data is unavailable at time of printing. Please contact Status Scientific Controls for updated
table.
Note:
Contact Status Scientific Controls for details regarding any sensor not listed.