EasyManua.ls Logo

STATUS SCIENTIFIC CONTROLS MCU3 - Analogue Output Configuration

Default Icon
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TD06/001
Issue:
6
Change Note:
1696
Page 4-11
4.1.2.5 3-Wire Connection (4-20mA Current Input and Separate Supply)
Connection Procedure
Disconnect power from the system.
Remove the grey protective cover from the appropriate input module
Configure the jumper switches as shown below (i.e. J6, J7, J10, J12, J13 and J14
switched ON, all others OFF)
Connect a voltmeter between pins 22 (+V) and 24 (0V).
Connect power to the system.
Adjust VR1 and VR2 until the correct voltage can be measured for the detector head
being used.
Disconnect power to the system.
Connect the detector head to the input module observing the above diagram.
Refit grey protective cover.
Connect power to the system and confirm that the detector head operates.
Perform software configuration (refer to ‘MCU Software Configuration and System
Calibration Manual’).
Perform system calibration. Refer to ‘MCU Software Configuration and System Calibration
Manual’ and detector head manufacturers instructions.
4.2 Analogue Output Configuration
In addition to providing the connections for various connector types, the Input Module can also
provide an analogue output. This output mimics the signal detected so that it may be used by
external equipment (e.g. chart recorders, data loggers) for a variety of purposes.
Note:
It is important to ensure that the Analogue Output is calibrated if in use (refer to ‘MCU Software
Configuration and System Calibration Manual’).
This configuration allows the connection of a detector head that provides
a 4-20mA current output, and requires a separate power source,
between 0-24V.
PL1
Terminals
0V
21
22
23
24
Current
Sensing
Circuit
Variable
Voltage
Output
0V
Voltage +
Input -
Current Output
J12
J11
J10
J9
J8
J7
J6
J14
J13

Table of Contents