7 Performance Test
© F. Stephan GmbH SA-103-0109V1.3-HAO-GB 29
7.7 Test of inspiration-and expiration valve
Fig. 8: Test of inspiration-and expiration valve
7.7.1 Basic setting
Circle system in operating mode
Remove mask from Y – piece
Connect the Y - piece with a test lung
Adjust fresh gas flow to 2 l/min
Limit pressure regulating valve to 35 mbar
7.7.2 Execution of the Test
Fig. 9: Test of inspiration- and expiration valve