10 Technical Data
36 SA-103-0109V1.3-HAO-GB © F. Stephan GmbH
10.3 Measuring range of the flowmeter tubes
O
2
(high precision) 0 to 2 l/min
(rough) 2 to 15 l/min
AIR 0 to 15 l/min
N
2
O 0 to 15 l/min
10.4 Accuracy of the flowmeter tubes
+/- 10 % of the respective terminal value of the scale.
In case of integrated high precision measuring range
+/- 10 % of the terminal value of this measuring range (under
standard conditions of 20
o
C and 1,013 bar)
10.5 O
2
-Failure alarm
Trigger pressure: 2,8 bar
Duration of audible alarm: 7 seconds
N
2
O - Blocking Trigger pressure: 2,0 bar
10.6 Technical Data Circle System
Volume of the complete circle
system with 2 absorbers and tubes: ca. 3 l
Weight: ca. 5,5 kg
Length: 330 mm
Height: 800 mm
Pressure range: -10 mbar to 100 mbar
Tolerance: +/- 5 % of the set value
Pressure regulating valve: 0 to 50 mbar
Pressure range: 0 to 50 mbar
Tolerance: +/- 5 mbar
Pressure gauge