Functional description
Button „Preoxy“
Press the „Preoxy“ button to administer
a pre-adjustable inspiratory oxygen
concentration for a certain preset
interval. These settings can be con-
figured in the „System Settings/Function“
menu. At the same time, the display of
the set oxygen concentration changes
to the pre-set „preoxy“ value. The alarm
limits for oxygen concentration are
adjusted automatically. The button
is green while pre-oxygenation is in
progress.
Button „Inspiration Hold“
Pressing this button during inspiration
will hold inspiration at the end
of the normal inspiration phase for
the duration for which the button is
pressed (maximum 15 s). Pressing
the button during expiration will
trigger a mandatory inspiration using
the set ventilation parameters.
Button „Day/Night“ Toggle
This button lets you toggle the
display, alarm indicator and
patient/function buttons
between day and night mode.