Section 17 — Appendix B - BEACH CHAIR and advanced hand control
82
56404485_C
STERIS 5095 General Surgical Table Operator Manual
7. Remove Beach Chair (refer to Instructions For Use, P150832-310).
IMPORTANT: Table leg section joints are still positioned downward.
8. Hold LEVEL button to level leg section joints.
9. Ensure leg section joints are level.
10. Reinstall table leg section.
FFiigguurree 1177--33.. EExxiittiinngg BBeeaacchh CChhaaiirr MMooddee
17.8 Level Table Function
Ensure surgical table is ready for operation as follows:
1. Ensure Head Section is properly installed.
2. Unplug table from facility power.
3. Verify table batteries are charged and table is ON.
4. If patient is on table, ensure patient is properly positioned.
5. Verify Advanced Hand Control is attached to a table side rail within easy reach of the operator.
6. Press LEVEL button to level table.