10
Setup and Interconnection
Stryker Endoscopy considers instructional training, or inservice, an integral
part of the PrecisionAC Camera. Your local Stryker Endoscopy sales
representative will perform at least one inservice at your convenience to help
set up your equipment and instruct you and your sta on its operation and
maintenance. To schedule an inservice, contact your local Stryker Endoscopy
representative after your equipment has arrived.
Setting up the PrecisionAC Camera involves three steps:
1. Setting up the console
2. Setting up the camera head
3. Setting up the coupler
• Always connect the console to an appropriate power
source, using a hospital-grade power cord. Loss of AC
power will cause the camera to shut down and the surgical
image to be lost.
• Only connect items to the PrecisionAC Camera that have
been specied for use with the camera system. Connecting
incompatible equipment may cause unexpected results.
• When the camera system is used with other equipment,
leakage currents may be additive. Ensure that all systems
are installed according to the requirements of IEC60601-1.
• Equipment which employs RF communications may aect
the normal function of the PrecisionAC Camera. When
choosing a location for the camera system, consult the
“Electromagnetic Compatibility” section to ensure proper
function.
• Always set up the console in a location that allows
adequate ventilation (airow) to the console. Insucient
ventilation may cause the console to overheat and shut
down.