39
User Maintenance
Maintenance, Inspection, and Testing
Inspect the device on a continual basis. If a problem is observed or
suspected, the device should be returned for repair.
Inspect all components for cleanliness. If uid or tissue buildup is present,
repeat the above cleaning and disinfection procedure.
Inspect the camera cable for cuts and breaks. Return any damaged camera
to Stryker for service.
Using Sterile Drapes
Using sterile drapes will ensure maximum longevity of your PrecisionAC
Camera. For best results, follow the instructions provided by the drape
manufacturer.
Storage
Never store the device in a non-ventilated, humid environment such as a
carrying case. This can present an infection control risk.
Replacing the Fuses
To avoid the risk of re, use only fuses of the value specied on
the fuse label located on the rear panel of the console.
1. Unplug the power cord from the wall outlet and remove the cord from
the transmitter console.
2. Unlatch the fuse holder above the AC inlet and remove it. (You may need
to press the tab on the fuse holder with a slender screwdriver to release
the latch.)
3. Replace the fuse with the same value and rating.
4. Reinstall the fuse holder until the tab snaps in place.