3
Warnings: SPY Mode
IMPORTANT SAFETY NOTICE - LASER RADIATION:
SPY mode controls a Class1M laser emitted from the L11 LED
Light Source with Advanced Imaging Modality (0220230300), also
referred to as the “L11 LED Light Source.
Use of controls or performance of procedures other than those
specied herein can result in hazardous laser radiation exposure
and can cause severe eye injury to the patient or user.
To avoid exposure to laser radiation, follow all warnings and
guidelines presented below and throughout this user manual.
1. Before using SPY mode, read and be familiar with all instructions and
warnings found in this user manual and the light source user manual.
2. Protect the camera system against unqualied use.
3. Wear eye protection as appropriate. Refer to any applicable regional
regulations or standards for personal protective equipment.
4. Do not manipulate tissue while Contrast mode (a SPY mode) is on.
5. When using SPY mode, do not view the light output with optical
instruments (for example, microscopes or magniers). Do not direct the
light output in SPY mode into an area where such instruments are likely
to be used.
6. Do not turn on SPY mode when the endoscope is outside of the patient’s
body.
7. When SPY mode is on, never look into the following apertures or direct
the light emitted from the apertures toward another person:
• the light cable connection on the light source (if the cable is not
attached)
• the end of the light cable (if the SafeLight™ adapter is attached)
• the endoscope tip
8. When SPY mode is on, never leave a SafeLight adapter attached to the
light cable without an endoscope attached. Laser radiation can continue
to emit from the adapter.
9. Disconnect the light cable from the light source only when the light
source is powered o or the light output is deactivated.