SSeeccuurriinngg tthhee ppaattiieenntt wwiitthh tthhee GG--rraatteedd rreessttrraaiinntt ssttrraappss
WWAARRNNIINNGG
Always use all restraint straps to secure the patient on the cot. An unrestrained patient may fall from the cot.
Do not attach restraint straps to the base tubes or cross tubes.
CCAAUUTTIIOONN -- Do not entangle the restraint straps in the base frame when you raise or lower the cot.
NNoottee -- Restraint straps are a Type BF applied part.
Secure restraint straps to the cot in the required attachment locations (Figure 10). Restraint strap attachment locations
should provide strong anchorage and proper restraint position. Do not allow restraint straps to interfere with equipment or
accessories. Buckle the restraint straps across the patient’s shoulders, waist, and legs. Buckle restraint straps when the cot
is not in use.
6086-209-001 Rev F.2 29 EN