KIPTRACK
52
Power Source:
Power saving:
Memory:
Measurement unit:
Linear range:
Pressure Range:
Heart Rate Range:
Measurement unit:
Accuracy of Pressure:
Accuracy of Heart rate:
Maximum Inflation pressure:
Systolic Measurement Range:
Diastolic Measurement Range:
External Output: RS232
IP Classification: IP21
Precision:
Accuracy:
KETONE warning:
Strip storage condition:
Monitor storage condition:
System operating condition:
Size of monitor w/o cuff:
+4% of reading
+15mg/dL
(0.83mmol/L)
when
4 C to 40 C (39.2 F to 104 F),
below 85% R.H.
10% to 95% RH
15% to 93% RH, 700 hPa to 1060 hPa
glucose
<
<
100mg/dL
5% (CV)
mg/dL or mmol/L
glucose value over 240 mg/dL (13.3 mmol/L)
20-600
mg/dL
(1.1
-
33.3
mmol/L)
(5.55mmol/L):
-
-
-
+15%
when
glucose
>
100mg/dL
(5.55
mmol/L)
-
+3mmHg or +2% of reading
-
-
mmHg or Kpa
40-199 beats per minutes
0-300 mmHg
300 mmHg
60 mmHg - 255 mmHg
30 mmHg - 195 mmHg
o
o
o
o
5 C to 40 C (41 F to 104 F),
o
o
o
o
352 measurement results with date & time
136mm (L) x 92 mm (W) x 56 mm (H)
Four 1.5V AA size alkaline batteries
Automatic power off if idle for 3 minutes
-25 C to 70 C (-13 F to 158 F),
o
o
o
o