EasyManua.ls Logo

TanGO TX1 - Page 51

Default Icon
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
Warnings persist until the event is resolved. In some cases, an unresolved warning will become more
urgent in frequency. For example, a low-battery warning that is not resolved will change to alarm status
indicating a critical low-battery condition.
As with alarm events, warning events are distinguished from one another on the instrument display screen
using different symbols. Table 5.2 lists the symbols used for different warning events.
Table 5.2 Warning events
Symbol
Warning event
Description
Calibration due
a
The instrument requires calibration.
Bump test due
a
The instrument requires a bump test.
Dock due
a
The instrument requires docking.
Acknowledgeable
Gas-alert
a
A detected gas concentration may be approaching alarm levels. To turn off
the warning signals, press and hold .
Sensor 1
Single sensor data failure
Sensor 1 or Sensor 2 is not installed or is in data failure.
Sensor 2
Critical Error
A critical error has occurred; a three-digit number code will be displayed.
Low-battery
The instrument’s battery is low.
a
Signals (visual, audible, and vibration) vary based on instrument settings.
Table 5.3 Warning and failure screens
Warnings
Calibration due
a
Bump test due
a
Dock due
a
Gas-alert (If enabled)
Low-battery warning
b
One sensor Installed
Failures
Single sensor calibration
failure
Single sensor bump failure
with calibration overdue
Single sensor zero failure
Single sensor data failure
a
Signals (visual, audible, and vibration) vary based on instrument settings.