EasyManua.ls Logo

TanGO TX1 - Product Overview

Default Icon
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2
Table 1.1 Certifications
Directive or CB
Area Classifications
Approved Temperature Range
China CPA
Metrology Pattern Approval
-20 °C ≤ Ta ≤ +50 °C (-4 °F ≤ Ta ≤ +122 °F)
a
The TANGO TX1 complies with relevant provisions of European ATEX directive 2014/34/EU and EMC directive 2014/30/EU.
b
The TANGO TX1 complies with relevant harmonized standards and legislation provisions for UK SI 2016 No. 1107, No. 1091, UK SI 2012 No. 3032.
Note: See also Marking Requirements for ATEX and IECEx in the Appendix.
Table 1.2 Warnings and cautionary statements
the unit. Failure to perform certain procedures or note certain conditions may impair the performance of this
product.
For safety reasons, this equipment must be operated and serviced by qualified personnel only.
Substitution of components may impair intrinsic safety and may cause an unsafe condition.
measure gas concentrations accurately. When this occurs, readings may appear lower than the actual gas
concentration. Keep sensor openings clean, dry, and properly exposed to the ambient air.
Obstructed, contaminated, or damaged sensor water barriers (or their gaskets) can inhibit the unit’s ability to
measure gas concentrations accurately. When this occurs, readings may appear lower than the actual gas
concentration. Replace the sensor water barriers and gaskets as needed (see “Service” for instructions).
use in oxygen-enriched atmospheres.
Sudden changes in atmospheric pressure, temperature, or humidity may cause temporary fluctuations in gas
readings.
following each of these incidences: the unit falls, is dropped, or experiences another significant impact; is
exposed to water; fails a bump test; or has been repeatedly exposed to an over-range (positive or negative) gas
concentration. Calibration is also recommended after the installation of a new (or replacement) sensor.
understand the potential hazards associated with calibration gas, please consult the Safety Data Sheets (SDS)
available on the Industrial Scientific website.
http://www.indsci.com/en/explore/calibration-gas-and-reference-
chart
Table 4.1 for instructions on how to set the time and date.
Product Overview
The Tango TX1 is a portable, long-life, single-gas monitor (instrument) for personal protection. It is a
diffusion instrument for use in detecting and measuring gas present in open space. Based on the