EasyManua.ls Logo

tecan NanoQuant Plate - Page 20

tecan NanoQuant Plate
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 Quick Guide NanoQuant Plate
TM
No.30035094 Rev No. 1.4 2013-06
Typical performance values Infinite 200 PRO
Wavelength absorbance
Wavelength accuracy < ± 0.5 nm for λ > 315 nm
< ± 0.3 nm for λ 315 nm
Wavelength reproducibility < ± 0.5 nm for λ > 315 nm
< ± 0.3 nm for λ 315 nm
Absorbance bandwidth Infinite M200 PRO < 5 nm for λ ≤ 315 nm
< 9 nm for λ > 315 nm
Absorbance bandwidth Infinite F200 PRO Depending on filters used
Measurement absorbance
Detection limit (DNA concentration) 1 ng/μl dsDNA
Reproducibility of one sample (50 μg/ml) < 1 % CV
Ratio 260/280 (50 μg/ml) ± 0.07
Precision @ 260 nm < 0.2 %
Accuracy @ 260 nm < 0.5 %
Figure 13: Technical drawing of the NanoQuant Plate (dimensions in mm)

Related product manuals