3. NanoQuant Plate
3.1 Parameters
Physical
Optics 16 quartz lenses (one per sample)
Quartz lens Optical path length: 0.5 mm diameter: 2.2 mm
Parallel measurement 16 sample positions (2 rows of 8)
Sample volume Min. 2 μl
Dimensions plate
Plate width: 85.4 mm Plate height: 14.6 mm Plate length:
127.8 mm Weight: ~ 160 g
Wavelength settings
Infinite M200 PRO
260 nm (5 nm bandwidth), 280 nm (5 nm bandwidth),
310 nm (5 nm bandwidth) reference
Infinite M1000
260 nm (5 nm bandwidth), 280 nm (5 nm bandwidth),
310 nm (5 nm bandwidth) reference
Infinite F200 PRO, Infinite F500
260 nm (5 nm bandwidth), 280 nm (3 nm bandwidth),
340 nm (10 nm bandwidth) reference
Measurement time of whole plate
Quantify Nucleic Acid Infinite 200 PRO: 1.15 minute for 16 samples
Labeling Efficiency Infinite 200 PRO: 2 minutes for 16 samples