EasyManua.ls Logo

Tecsis EZE30 - Page 11

Tecsis EZE30
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DISPLAY THE INPUT SIGNAL IN mV/V
Use this function to view the raw signal coming from the load cells into the EZE30. The display shows the
load cell signal directly in mV/V. This can be useful when either commissioning or fault finding.
0
T
1
T
T
T
2
0
T
1
CALIBRATE THE SPAN FROM THE LOAD CELL mV/V READING
If you are calibrating the weighing system without weights, use this menu point to enter the mV/V value for
the weighing system. The weighing system mV/V value can be derived from the load cell test data. Normally
a test data certificate is provided with each load cell stating amongst other things, the Output at Rated Load
(ORL) in mV/V. In single load cell applications the weighing system mV/V value will be the load cell ORL
value. In applications using more than one load cell, the weighing system mV/V value will be the average of
the individual load cells ORLs. Normally load cells used in multi-cell applications have matched or
rationalized ORLs typically 2 or 3 mV/V for example. So if you have 4 load cells in your system, each 1000
kg capacity with ORLs of 2 mV/V then the weighing system capacity will be 4000 kg at 2 mV/V.
Use the UP/DOWN & MOVE RIGHT keys to enter the weighing system ORL reading in mV/V. The reading
should be in the format x.xxxx e.g 2.0000 if the ORL is 2 mV/V exactly. The corresponding weighing system
capacity should be entered in menu point 2.1
0
T
1
T
T
T
2
0
T
1
3
4
SET UP THE DISPLAY - See Next Page
SET UP AND CALIBRATE SPAN /DISPLAY INPUT SIGNAL IN mV/V (Continued)
3
4
Page11