Performance
Check
Procedure
NOTE
The
PG
506
must
be
removed
from
the
TM
power
supply
to
make
the
change
to
de
output
from
the
generator.
Turn
the
power
off
before
removing
or
inserting
any
plug-in
from
the
TM
power
supply
Connect
the
callbration
generator
Std
Ampl
output
to
the
CH
1
input
via
a
50
2
precision
BNC
coaxial
cable.
Set
calibration
generator
for
Std
Ampl
output
of
50
mV
dc.
Set:
CH
1
VOLTS/DIV
10
mV
CH
1
Input
COUPLING
Dc
CHECK—the
readout
is
between
49.0
mV
and
51.0
mV.
Set
CH
1
VOLTS/DIV
to
0.1
V.
Set
calibration
generator
for
Std
Ampl
output
of
0.5
V.
CHECK—the
readout
Is
between
0.495
V
and
0.505
V.
Set
CH
1
VOLTS/DIV
to
1
V.
Set
calibration
generator
for
Std
Ampl
output
of
5
V.
CHECK—the
readout
Is
between
4.95
V
and
5.05
V.
Disconnect
Std
Ampl
signal
from
the
CH
1
input.
DC
Volts
Normal
Mode
Rejection
Ratio
Set
SEC/DIV
to
5
ms.
Connect
function
generator
(FG
502)
output
to
the
CH
1
Input
via
a
50
©
BNC
coaxial
cable
and
a
BNC
coupling
capacitor.
Set
function
generator
for
a
six—division
sIne-wave
display
amplitude
at
50
Hz
(with
CH
1
VOLTS/DIV
at
1
V).
Set
CH
1
VOLTS/DIV
to
0.2
V.