Loading...
Performance
Check
Procedure
CH
1/CH
2
VOLTMETER
Equipment
Required
(See
Table
7-1)
Calibration
generator
Leveled
sine-wave
generator
Function
generator
50
©
BNC
coaxial
cable
50
Q
BNC
termination
BNC
coupling
capacitor
DC
Volts
Accuracy
Set:
READOUT
(intensity)
For
a
viewable
readout
A
INTEN
For
a
viewable
trace
Vertical
MODE
CH1
CH
1
VOLTS/DIV
50
mV
CH
2
INVERT
Off
SCOPE
BW
Off
CH
1
Input
COUPLING
GND
Horizontal
MODE
A
A
SEC/DIV
ims
A/B
SELECT
A
Trigger
Trigger
MODE
AUTO
LEVEL
Trigger
CPLG
oc
Trigger
SOURCE
VERT
Trigger
SLOPE
_~
(positive-
going)
Trigger
HOLDOFF
MIN
J
FOCUS
For
best
defined
display
-™
Horizontal
POSITION
12
o'clock
\
ee}
b.
Press
the
VOLTMETER
button
and
then
press
DC
on
the
VOLTMETER
menu.
7
|
c.
CHECK—ground
readout
Is
0.0
mV
+1.2
mV.
d.
Set
calibration
generator
(PG
506)
internal
Square
Wave/DC
switch
to
-
as
2247A
Operators
7-51