3. Click to set the Horizontal Delay Mode button to On.
4. Select the trigger delay time with the Trig Delay control.
5. Use the Horiz Delay control to delay the acquisition relative to the B event.
6. Adjust the Ref Point and B Trig Level controls as necessary.
Behavior. Use this trigger sequence to trigger the instrument on the B trigger event after the A event.
When you select Trig After Time, the instrument looks for the trigger conditions specified by the A event. After the A event arms
the trigger system, the instrument delays the trigger for the period of time specified by the Trig Delay control. After the trigger
delay time has expired, the instrument looks for the B trigger event. When the instrument detects the B event, it delays the
acquisition for the period of time specified by the Horiz Delay.
What do you want to do next?
Learn more about Horizontal Delay.
Learn more about other trigger types.
Return to the Trigger Setup control window.
Trigger setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 449

Table of Contents

Other manuals for Tektronix DPO5104B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5104B and is the answer not in the manual?

Tektronix DPO5104B Specifications

General IconGeneral
BrandTektronix
ModelDPO5104B
CategoryTest Equipment
LanguageEnglish

Related product manuals