4. Select the trigger delay time with the Trig Delay control.
5. Adjust the Horiz Position and B Trig Level controls as necessary.
Behavior. When you select Trig After Time, the instrument looks for the trigger conditions specified by the A event. After the A
event arms the trigger system, the instrument triggers on the next B event that occurs after the time specified by the Trig Delay
control.
What do you want to do next?
Learn more about other trigger types.
Return to the Trigger Setup control window.
Trigger setups