EasyManua.ls Logo

Tektronix MSO4000B Series - Page 73

Tektronix MSO4000B Series
87 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verication
c. Set the Termina
tion (input impedance) to 1M.
d. Repeat steps 9 through 13.
15. Repeat the pro cedure for all remaining channels as follows:
a. Push the front-panel button to deselect the channel that you have already
tested.
b. Move the DC voltage source connection to the next channel input to be
tested.
c. Starting from step 6, repeat the procedure for each channel.
MSO4000B and DPO4000B Series Specications and Performance Verication 65

Table of Contents

Other manuals for Tektronix MSO4000B Series

Related product manuals