EasyManua.ls Logo

Tektronix P6137 - Page 14

Tektronix P6137
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
P6137 10X Passive Probe Instructions
5
Probe Adjustment/Maintenance
WARNING. The following servicing instructions are for use by qualified
personnel only. To avoid electrical shock, do not perform any probe maintenance
while the probe is connected to a signal source.
Probe Compensation
Due to variations in oscilloscope input characteristics, probe low-frequency
compensation should be checked and adjusted after moving the probe from one
input to another.
To adjust low-frequency compensation, apply a 1 kHz square-wave signal (such
as an oscilloscope calibrator output) to the probe tip. Using a low-reactance
alignment tool, adjust the probe’s compensation capacitor through the hole in the
compensation box to obtain the squarest waveform at the front corner.
High-frequency compensation seldom requires adjustment; however, if the probe
has excessive high-frequency aberrations or insufficient bandwidth, then adjust
the high-frequency compensation.
Typical test equipment required includes a Tektronix PG506A pulse generator
(rise time of 1 ns) & Tunnel Diode Pulser (Tektronix part number 067-0681-01)
with pulse rise time of 125 ps. Use a 50 W terminated BNC-to-compact probe
tip adapter (such as Tektronix part number 013-0227-00) to connect the probe to
the equipment.
To adjust high-frequency compensation, first remove the plastic cover (see the
parts replacement procedures). Next, adjust R1 and R3 (shown in Figure 6) for
best long-term flatness. Finally, adjust R2 and C1 for best short-term flatness.
Due to interaction between controls, the adjustment procedure may need to be
repeated. Refer to Figure 7 for additional information about compensation
procedures.

Related product manuals