Performance Tests
Performance Verification
4Ć28
Third read the
results of the
DDC
measurement here.
Second align the second
cursor to the DC test
level that you input.
First align a cursor to the
DC baseline (no input).
Figure 4Ć6:ăMeasurement of the DC Accuracy for Delta Measurements
Test all channels: ! " ! !" "
Reestablish the initial test hookup setup:
Hook up the testĆsignal source:
H ! ! "!"! ! ! ! #!
H # ! ! ! CH 1
Initialize the oscilloscope:
H # SETUP
H ! $" "!! Recall Factory Setup
H ! $" "!! OK Confirm Factory Init
Modify the default settings:
H SHIFT ! ACQUIRE MENU
H ! $" "!! Mode ! ! $"
"!! Average 16
H DISPLAY
H ! $" "!! Graticule ! ! $"
"!! Frame