Performance Tests
4-40
TDS 420A, TDS 430A & TDS 460A Service Manual
b. Select the Delayed trigger system:
H Press SHIFT; then press the front-panel button DELAYED TRIG.
H Press the main-menu button Level.
c. Measure the test signal: Press the side-menu
button SET TO 50%.
Read the measurement results in the side menu below the label Level.
d. Check against limits: Do the following subparts in the order listed.
H Subtract the trigger level readout from the DC calibration generator
setting. The result is the trigger level accuracy.
H CHECK that the trigger level is accurate to within 18 mV.
H Press TRIGGER MENU. Press the main-menu button Slope; then
press the side-menu button for negative slope. (See icon at left.)
H Subtract the trigger level readout from the DC calibration generator
setting. The result is the trigger level accuracy.
H CHECK that the trigger level is accurate to within 18 mV.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Then disconnect the cable from the generator output at the input
connector of CH 1.
Equipment
Required
One sine wave generator (Item 11)
One precision 50 coaxial cable (Item 5)
One 10X attenuator (Item 1)
One 5X attenuator (Item 2)
Prerequisites See page 4--15.
1. Install the test hookup and preset the oscilloscope controls:
a. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the main-menu button Recall Factory Setup.
H Press the side-menu button OK Confirm Factory Init.
Sensitivity, Edge Trigger,
DC Coupled