Performance Tests
4-44
TDS 420A, TDS 430A & TDS 460A Service Manual
b. Set the Main and Delayed Horizontal Scales:
H Set the horizontal SCALE to5nsfortheM (Main) time base.
H Press the side-menu button Delayed Only.
H Set the horizontal SCALE to2nsfortheD (Delayed) time base.
Press the side-menu button Main Only.
c. Display the test signal:
H Set the generator frequency to 350 MHz (TDS 420A) or 500 MHz
(TDS 430A and TDS 460A).
H Set the test signal amplitude for about five divisions on screen. Now
fine adjust the generator output until the CH 1 Amplitude readout
indicates the amplitude is 500 mV. (Readout may fluctuate around
500 mV.)
H Disconnect the sine wave at CH 1 and reconnect it to CH 1 through
a 5X attenuator.
d. Repeat step 2, substeps b and c only.
4. Confirm that the Main and Delayed trigger systems couple trigger signals
from all channels: Doing the procedure Check Analog Bandwidth, which
begins on page 4--25, checks coupling. If you have not done that procedure,
do so after finishing this procedure. See the following note.
NOTE. Steps 1 through 3 confirmed trigger sensitivity for the Main and Delayed
triggering systems using the CH 1 input. Doing the procedure Check Analog
Bandwidth ensures that trigger signals are coupled from all channels.
5. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of the channel last tested.
Equipment
Required
One sine wave generator (Item 11)
Two precision 50 coaxial cables (Item 5)
One BNC T connector (Item 8)
Prerequisites See page 4--15.
1. Install the test hookup and preset the oscilloscope controls:
a. Initialize the oscilloscope:
H Press save/recall SETUP.
Maximum Input