Performance Tests
4-28
TDS 420A, TDS 430A & TDS 460A Service Manual
Table 4- 4: Analog Bandwidth (TDS 430A and TDS 460A)
Vertical Scale Reference Amplitude Horizontal Scale Test Frequency Limits
100 mV 600 mV (6 divisions) 2ns 400 MHz 424 mV
1V
4
2 V (2 divisions) 2ns 400 MHz 1.414 V
500 mV 2 V (4 divisions) 2ns 400 MHz 1.414 V
200 mV 1.2 V (6 divisions) 2ns 400 MHz 848 mV
50 mV 300 mV (6 divisions) 2ns 400 MHz 212 mV
20 mV 120 mV (6 divisions) 2ns 400 MHz 84 mV
10 mV 60 mV (6 divisions) 2ns 400 MHz 42 mV
5mV 30 mV (6 divisions) 2ns 400 MHz 21 mV
2mV 12 mV (6 divisions) 2ns 250 MHz 8.4 mV
1mV 6 mV (6 divisions) 2ns 100 MHz 4.2 mV
4
If the oscilloscope does not trigger in step 2e, change the trigger coupling to DC for this Vertical Scale setting.
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
H Press MEASURE; then press the main-menu button Select
Measurement for CHx.
H Now press the side menu button more until the menu label Pk-Pk
appears in the side menu (its icon is shown at the left). Press the
side-menu button Pk-Pk.
H Press CLEAR MENU.
H Set the sine wave generator output (if necessary, use 10X attenua-
tors) so the CHx Pk-Pk readout equals the reference amplitude in
Table 4--3 (TDS 420A) or Table 4--4 (TDS 430A and TDS 460A)
that corresponds to the vertical scale set in substep d.
H Press the front-panel button SET LEVEL TO 50% as necessary to
trigger a stable display.
f. Measure the test signal:
H Increase the frequency of the generator output (leveled output) to the
test frequency in Table 4--3 (TDS 420A) or Table 4--4 (TDS 430A
and TDS 460A) that corresponds to the vertical scale set in
substep d.
H Set the horizontal SCALE to 5 ns (TDS 420A) or 2 ns (TDS 430A
and TDS 460A).
H Press SET LEVEL TO 50% as necessary to trigger the display.

Table of Contents

Other manuals for Tektronix TDS 420A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 420A and is the answer not in the manual?

Tektronix TDS 420A Specifications

General IconGeneral
BrandTektronix
ModelTDS 420A
CategoryTest Equipment
LanguageEnglish

Related product manuals