Command Descriptions
2−46
TDS200/1000/2000 Series Oscilloscope Programmer Manual
BUSY?
might return 1, indicating that the oscilloscope is busy.
*OPC, *WAI
*CAL? (Query Only)
Performs an internal self-calibration and returns its status. This is
equivalent to selecting the Do Self Cal option in the Utility menu.
Although *CAL? is a query command, it does perform an action.
NOTE. The self-calibration can take several minutes to complete.
During this time, the oscilloscope does not execute any commands.
Disconnect all signals from the oscilloscope before performing an
internal self-calibration.
Calibration and Diagnostic
*CAL?
0 indicates that the self-calibration completed without any errors
detected.
Any value other than zero indicates that the self-calibration did not
complete successfully or completed with errors.
*CAL?
performs a self-calibration and might return 0 to indicate that it
completed successfully.
Examples
Related Commands
Group
Syntax
Returns
Examples

Table of Contents

Other manuals for Tektronix TDS200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS200 Series and is the answer not in the manual?

Tektronix TDS200 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals