Performance Tests
4-50
TDS7104 & TDS7054 Service Manual
d. Save CH 3 waveform:
H Move the power divider from CH 2 to CH 3, so that CH 1 and
CH 3 are driven. Press the Vertical CH 2 and CH 3 buttons. Select
the Ref 3 tab and touch the Ch Channel 3 button. Touch the Save
Wfm to Ref3 Save button.
e. Display all test signals:
H Press the CH 3 Vertical button to remove CH 3 from the display.
H Display the live waveform. Move the power divider from CH 3 to
CH 4, so that CH 1 and CH 4 are driven. Press the Vertical CH 4
button to display. See Figure 4--22 on page 4--51.
H Display the reference waveforms. To do this, touch the Ref 3
Display Off button to toggle it to On and display the reference.
Select the Ref 2 tab and touch the Display Off button to toggle it to
On. You may notice their overlapping waveform handle icons. See
Figure 4--22 on page 4--51.
f. Measure the test signal:
H Locate the time reference points for these waveforms. Do this by
first identifying the point where the rising edge of the left-most
waveform crosses the center horizontal graticule line. Next, note the
corresponding time reference point for the right-most waveform. See
Figure 4--22 on page 4--51.
H Press CURSORS and select the V Bars Cursors Type.
H Touch the Close button.

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals