EasyManuals Logo

Tektronix TDS6000 Series Service Manual

Tektronix TDS6000 Series
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Performance Tests
TDS7104 & TDS7054 Service Manual
4-79
H Touch the Save Wfm to Ref1 Save button to save the probe
compensation signal in reference 1.
H Disconnect the cable from CH 1 and the probe compensation
connector.
H Touch the Display button to toggle it to on to displayed the stored
signal.
d. Hook up the DC standard source:
H Set the output of a DC calibration generator to off or 0 volts.
H Connect the output of a DC calibration generator through a
dual-banana connector followed by a 50 precision coaxial cable to
one side of a BNC T connector. See Figure 4--37.
H Connect the Sense input of the generator through a second dual-ba-
nana connector followed by a 50 precision coaxial cable to the
other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 4--37.
TDS7000 oscilloscope
DC calibrator
50 coaxial cables
Dual banana to
BNC adapters
BNC T
connector
Figure 4- 37: Subsequent test hookup
e. Measure amplitude of the probe compensation signal:
H From the tool bar, touch Horiz and select the Acquisition tab.
H Touch Average and set the number of averages to 16 using the
keypad or the multipurpose knob.
H Adjust the output of the DC calibration generator until it precisely
overlaps the top (upper) level of the stored probe compensation
signal. (This value will be near 1000 mV).

Table of Contents

Other manuals for Tektronix TDS6000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000 Series and is the answer not in the manual?

Tektronix TDS6000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals