Brief Procedures
TDS7104 & TDS7054 Service Manual
4-15
d. Locate and then double click in the dialog box on the setup file that you
previously stored.
e. Verify that the oscilloscope retrieved the saved setup from the disk. Do
this by noticing the horizontal SCALE is again 1 ms and the waveform
shows ten cycles just as it did when you saved the setup.
9. Remove the test hookup:
a. Disconnect the BNC cable from the channel input and the probe
compensation output.
b. Remove the floppy disk from the floppy disk drive.