How to Test the DC Voltages on 3000TA Rear Panel Board?
Scope:
Power Module used in 3000TA is compact & maintenance free module,
which normally doesn’t give any kind of maintenance or service. If,
however, it is observed that display is missing due to any voltage problem.
It can measured as follows.
Parts:
Multimeter
Tools:
Screw Driver set.
Procedure:
Remove power to the instrument by unplugging the power cord at the
power source.
Follow the below procedure for removing the rear panel assembly.
To detach the rear panel, remove only the 14 screws marked with an X.
Position the rear panel assembly properly & apply the power supply to the
instrument.
Locate the Power supply module on the Rear Panel Board. Kept the
Multimeter in DC Voltage range (0-20 or 40 VDC range).