EasyManua.ls Logo

Teledyne 326 - Maintenance and Troubleshooting; Routine Maintenance and Cell Life Management; Special Shut-Down Precautions for A-3 Cell

Teledyne 326
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Teledyne Analytical Instruments
MaintenanceMaintenance
MaintenanceMaintenance
Maintenance
Model 326/327/328Model 326/327/328
Model 326/327/328Model 326/327/328
Model 326/327/328
5-1
Maintenance & Maintenance &
Maintenance & Maintenance &
Maintenance &
TT
TT
T
rr
rr
r
ouboub
ouboub
oub
leshootingleshooting
leshootingleshooting
leshooting
5.1 Maintenance5.1 Maintenance
5.1 Maintenance5.1 Maintenance
5.1 Maintenance
5.1.1 Routine Maintenance5.1.1 Routine Maintenance
5.1.1 Routine Maintenance5.1.1 Routine Maintenance
5.1.1 Routine Maintenance
Most 326 series analyzers contain no moving parts (although some
contain an analog meter movement, solenoid, or relay contacts), and there-
fore, periodic service is not required. The sample flow check and calibration
outlined in Section 4 should be adequate to keep the analyzer functioning
throughout the life span of a given Micro-Fuel Cell.
Special designs may require sample systems with filter elements, valves,
etc., and those components will require inspection or replacement as needed
to keep the instrument in service. A periodic inspection schedule (once per
week, once per month, etc.) will be sufficient to determine problems before
they become serious.
5.1.2 Special Shut-Down Precaution for Class A-3 Cell5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.2 Special Shut-Down Precaution for Class A-3 Cell5.1.2 Special Shut-Down Precaution for Class A-3 Cell
5.1.2 Special Shut-Down Precaution for Class A-3 Cell
When the analyzer is equipped with a Class A-3 cell, it must be pro-
tected from prolonged exposure (more than 4 hours) to CO
2
-free atmo-
spheres (such as air). This can be accomplished by trapping CO
2
-containing
sample (flue gas) by closing the input manifold valves and sealing the vent
connection with a valve, water trap or clamped-off plastic tubing.
5.1.3 Cell Replacement5.1.3 Cell Replacement
5.1.3 Cell Replacement5.1.3 Cell Replacement
5.1.3 Cell Replacement
When the Micro-Fuel Cell nears the end of its useful life, sensitivity will
decline very rapidly. The initial response to this phenomenon is to
recalibrate the analyzer. If many turns of span control are required, or if the
control does not have sufficient range to recalibrate the analyzer, a new
Micro-Fuel Cell is required.