The Quick Start Guide describes the steps to successfully start and
calibrate the instrument. This guide covers the AX/MX 300 series
of analyzers and monitors. Before placing the instrument in service
please read the users manual that accompanies the device.
Components
The AX/MX 300 Portable Oxygen Monitor includes:
Sensor & Diverter
Sensor Cable
T-Adapter
V mount Adapter
AA Batteries
Instruction Manual
Set Up
Set up is simple:
1. Connect the coiled sensor cable to the
monitor and tighten the capture nut.
2. Connect the other end to the sensor
and tighten the capture nut.
3. Install 3 AA alkaline batteries.
Calibration
Note: The AX300 and MX300 can only be calibrated
using 100% oxygen or room air (20.9%). Improper
calibration or use of other gas concentrations will
activate the
SENSOR indicator.
Battery Installation
1. Hold the instrument face down.
2. Using a coin. Gently twist to open
the hinged battery compartment door
latch.
3. Lift up door near latch and pull out.
4. Install 3 AA alkaline batteries
observing proper polarity.
Note: Depressing keys while installing
battries may activate an error code.To
reset remove and replace battery.
Quick Start Guide for Model AX300 Portable Oxygen Anal
yzer and
MX300 Portable Oxygen Monitor

Other manuals for Teledyne AX300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Teledyne AX300 and is the answer not in the manual?

Teledyne AX300 Specifications

General IconGeneral
Response Time< 60 seconds to 95%
Operating Temperature5-40°C (41-104°F)
Range0-100 ppm
Measurement Range0-1 ppm, 0-10 ppm, 0-100 ppm (user selectable)
Output4-20 mA

Related product manuals