EasyManua.ls Logo

TELEVAC MM200 - Page 132

TELEVAC MM200
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 131 of 160
TELEVAC
PROCEDURE FOR ON-SITE VACUUM CALIBRATION OF TELEVAC MODULAR MM200 INSTRUMENTS
1. Install the master gauge tubes on the vacuum chamber in a location, which will allow the
master gauge tubes and the vacuum chamber gauge tubes to see the same pressure under
static test conditions. To achieve this, we recommend that a "T" be installed on each vacuum
chamber. (See figures).
2. Pump the chamber down to a pressure of less than 10
- 4
Torr.
3. On the main chamber thermocouple gauge (usually TC#1), adjust the "zero adjust" ("heater
adjust") potentiometer to bring its reading into agreement with the master gauge TC#1. It
should read at " 0.0" ("red line") on the MM200 instrument.
4. Close the main valve. Slowly bleed dry nitrogen into the chamber until a pressure of 1 micron
is observed on the master thermocouple gauge. Observe the gauges for a minimum of 2
minutes. Leak up should be minimal before proceeding. If not, determine the source of the
leak and correct it.
5. Check the thermocouple readings at 1, 10 & 150 microns. Under the conditions of field test, at
the vacuum chamber, the following tolerances should be observed:
1 + 1 micron
10 + 2 microns
150 + 30 microns
Record the data. Readjust, if required, and record new data.
6. With the pressure at 1 micron (10
-3
Torr ), observe that the readings on the cold cathode gauge
of the vacuum chamber instrument, fall between 8 x 10
-4
and 3 x 10
-3
Torr.
Minor adjustment of the gauge can be made to show agreement with the thermocouple gauge
at 1 micron.In order to do this, adjust the cold cathode " 10
-3
" potentiometer to give a meter
reading of 1 x 10
-3
Torr.
Record data.
7. Place calibration sticker on instrument.
8. Remove the master gauges from the vacuum chamber and replace the plugs in each
calibration "T".

Table of Contents