INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 69 of 160
TELEVAC
NEW COMMANDS FOR MM200 - FOR SOFTWARE VERSION 2.31 OR HIGHER
EBH Cold Cathode station 6 to BOTH mode.
EBL Cold Cathode station 5 to BOTH mode
EHH Cold Cathode station 6 to AUTO mode.
EHL Cold Cathode station 5 to AUTO mode.
ESH Cold Cathode station 6 to SELF mode.
ESL Cold Cathode station 5 to SELF mode.
FFH Turn filament off for station 6, 3F gauge. The higher position of two possible 3F gauges.
FFL Turn filament off for station 5, 3F gauge. The lower position of two possible 3F gauges.
FH1 Switch to filament number 1 for station 6, 3F gauge.
FH2 Switch to filament number 2 for station 6, 3F gauge.
FL1 Switch to filament number 1 for station 5, 3F gauge.
FL2 Switch to filament number 2 for station 5, 3F gauge.
FNH Turn the filament on for station 6, 3F gauge. –not working yet--
FNL Turn the filament on for station 5, 3F gauge.
GFH Turn degas off for station 6, 3F gauge.
GFL Turn degas off for station 5, 3F gauge.
GNH Turn degas on for station 6, 3F gauge.
GNL Turn degas on for station 5, 3F gauge.
RO This command outputs the voltage being received from each of the sensor modules plus stations 8 and 9.
Stations 8 and 9 are actually the filament drive being supplied to stations 5 and 6 when 3F gauge(s) are
installed. This command can be used place of the SD command that is used with 3D or 3E gauges. Use
CA command or X to stop.
COMMANDS THAT HAVE BEEN UPDATED FOR 3F MODULES
GTnnn Set degas time to nnn minutes. (must include lead zeroes) When time is set to zero, the
degassing is untimed and must be ended manually through setup. In this case the two 3Fs
can be individually timed (manually). When any other value is set both station 5 and/or
station 6 will stop degassing when the time runs down to zero.
Rx Read station x
Sx Send station x process type.
ST Send status.
OTHER
SD Send filament drive, use RO. Filament drive for station five is displayed as station 8, filament drive for
station 6 is displayed as station 9. Use CA command or X to stop.