INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 47 of 160
TELEVAC
7B2 - 0-10V OUTPUT OPTIONS
RS 232
CODE
R7BM
R7BN
R7BE
R7BL
Volts
Volts
Volts
Volts
1X10-3
10.0
10.000
10.00000
10.00000
9X10-4
9.8
9.000
9.88561
9.84747
8X10-4
9.6
8.000
9.75772
9.67697
7X10-4
9.4
7.000
9.61275
9.48366
6X10-4
9.2
6.000
9.44538
9.26050
5X10-4
9.0
5.000
9.24743
8.99657
4X10-4
8.8
4.000
9.00515
8.67353
3X10-4
8.6
3.000
8.69280
8.25707
2X10-4
8.4
2.000
8.25257
7.67010
1X10-4
8.2
1.000
7.50000
6.66667
9X10-5
7.8
0.900
7.38561
6.51414
8X10-5
7.6
0.800
7.25772
6.34363
7X10-5
7.4
0.700
7.11275
6.15033
6X10-5
7.2
0.600
6.94538
5.92717
5X10-5
7.0
0.500
6.74743
5.66323
4X10-5
6.8
0.400
6.50515
5.34020
3X10-5
6.6
0.300
6.19280
4.92374
2X10-5
6.4
0.200
5.75257
4.33677
1X10-5
6.2
0.100
5.00000
3.33333
9X10-6
5.8
0.090
4.88561
3.18081
8X10-6
5.6
0.080
4.75772
3.01030
7X10-6
5.4
0.070
4.61275
2.81699
6X10-6
5.2
0.060
4.44538
2.59384
5X10-6
5.0
0.050
4.24743
2.32990
4X10-6
4.8
0.040
4.00515
2.00687
3X10-6
4.6
0.030
3.69280
1.59040
2X10-6
4.4
0.020
3.25257
1.00343
1X10-6
4.2
0.010
2.50000
0.00000
9X10-7
3.8
0.009
2.38561
8X10-7
3.6
0.008
2.25772
7X10-7
3.4
0.007
2.11275
6X10-7
3.2
0.006
1.94538
5X10-7
3.0
0.005
1.74743
4X10-7
2.8
0.004
1.50515
3X10-7
2.6
0.003
1.19280
2X10-7
2.4
0.002
0.75257
1X10-7
2.2
0.001
0.00000

Table of Contents