INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 78 of 160
TELEVAC
"SC<cr>" 2. Send all processor types. This causes the processor types for all ten stations
to be dumped. They are sent in a string of ten digits (or letters). The first digit or
letter corresponds to station #1, and so forth.
The digits or letter that corresponds to the sensor types are:
SENSOR TYPE
0 none installed
1 7F Extra wide range CC
2 3E E-Beam degas HC
3 2A Thermocouple
4 4A Convection gauge
5 1F 10 bar diaphragm
6 1E 1000 Torr diaphragm
7 3D I-sqrd-R degas HC
8 7B Standard CC
9 5A 1000 Torr Capacitance Diaphragm Gauge
A 7E Wide range CC
B 5D 1000 micron Capacitance Diaphragm Gauge
C 5B 100 Torr capacitance Diaphragm Gauge
D 5C 10 Torr Capacitance Diaphragm Gauge
E 5E 10 micron Capacitance Diaphragm Gauge
F 5F 10,000 Torr Capacitance Diaphragm Gauge or
other special option
A "0" means that there is no sensor assigned to this station.
If a hot cathode model 3D, or 3E is installed, then only five digits, or letters will
be sent, and the hot cathode gauge will always occupy station number five. If
a cold cathode is installed then only 9 digits will be sent.
"Sx<cr>" 3. Send processor type for a station. The modular unit responds with the sensor
type that is installed in station "x". For example if "S1<cr>" were sent, and
station number 1 contained a Model 2A thermocouple module, the response
would be "S1=2A<cr>". If station 1 were not occupied, the response would be
"S2=none<cr>". In the burst mode, the number or letter indicated in 2 above
will be sent.
"SV<cr>" 4. Fetch the modular unit software version. The modular unit responds with "Ver
n.nn<cr>", where n.nn is the version number.