INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 87 of 160
TELEVAC
"SSxFyyyyu<cr>" 11. Set the OFF set point of a particular relay. Where "x" is the relay number for
which it is desired to set the dropout (de-energize) point, and "nnnn" is the
setting. The response to this command is "A<cr>". See the chart below for
ranges of settings for various sensors.
"SSxNyyyyu<cr>" 12. Set the ON set point of a particular relay. Where "x" is the relay number, and
"nnnn" is the setting. See the chart below for ranges of settings for the
various sensors.
"SSxFy.y-n<cr>" 13. Set the OFF set point of a particular relay. Where "x" is the relay number for
which it is desired to set the dropout (de-energize) point, and "y.y-n" is the
exponetial setting. The response to this command is
"A<cr>". See the chart below for ranges of settings for various sensors.
"SSxNy.y-n<cr>" 14. Set the ON set point of a particular relay. Where "x" is the relay number, and
"y.y-n" is the exponential setting. See the chart below for ranges of settings
for the various sensors.