INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 96 of 160
COMMUNICATIONS (Can be stored)
AD<cr> Add RS485 response delay.
AT<cr> Add 50 millisecond command timeout.
BE<cr> Blank echo.
CT<cr> Clear 50 millisecond command timeout.
EAx<cr> Enter RS485 address code "x".
EE<cr> Enable echo.
G2<cr> Go to the RS422 mode NOT USED
G4<cr> Go to the RS485 mode.
PE<cr> Insert and test for even parity.
PF<cr> Turn off parity check and insertion.
PO<cr> Insert and test for odd parity.
RA<cr> Read RS485 address code byte.
RD<cr> Remove RS485 response delay.
RR<cr> Return to the RS232 mode.
SBll<cr> Set baud rate (See Table).
SE<cr> Store entries.
CONFIGURATION
AR<cr> Outputs relay configuration.
SC<cr> Send all stations configuration.
Sx<cr> Returns processor type installed in Station x.
SV<cr> Send software version.
DATA OUTPUT
Axxx<cr> Output each marked station value
every xxx cycles, where xxx is any value
between 1 and 255.
BO<cr> Output data in burst mode (See above).
CA<cr> Cancel the automatic output.`
Mx<cr> Mark station x, where x is any number between
1 and 9, or A for Station 10.
Rx<cr> One shot reading of station x.
Ux<cr> Unmark station #x from automatic output.
DISPLAY
DLx<cr> Station x is displayed on the left side.
DRx<cr> Station x is displayed on the right side.
D1E<cr> 1F gauges to be displayed in exponential form.
D1L<cr> 1F gauges to be displayed in linear form.
FA(cr) In conjunction with the "FI" command, will inhibit
all front panel function, except station select.
FE<cr> Enables certain front panel functions, unless
inhibited by the rear panel switch.