EasyManua.ls Logo

TELEVAC MM200 - Page 99

TELEVAC MM200
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:sales@televac.com web site:www.televac.com
MM-200_im REV M
Page 98 of 160
TELEVAC
R2AR<cr> Recorder output of 2As to be reversed linear to (1) Torr.
R2AT<cr> Recorder output of 2As to be 2A traditional.
R3DL<cr> Recorder output of 3D or 3E to be logarithmic.
R3DM<cr> Recorder output of 3D or 3E to be multi linear.
R3Dx<cr> Recorder output of 3D or 3E to be linear to 10-x Torr
where 8 => x >= 2 and 10V is at 10-x Torr.
R4AL<cr> Recorder output of 4As to be logarithmic.
R4AN<cr> Recorder output of 4As to be multi linear.
R4AT<cr> Recorder output of 4As to be linear to 10 Torr.
R4AU<cr> Recorder output of 4As to be linear to 1 Torr.
R5AL<cr> Recorder output of 5As & 5Bs to be logarithmic.
R5AN<cr> Recorder output of 5As & 5Bs to be linear.
R7BA<cr> Recorder output of 7Bs to be 7AC traditional.
R7BE<cr> Recorder output of 7Bs to be extended log.
R7BL<cr> Recorder output of 7Bs to be logarithmic.
R7BM<cr> Recorder output of 7Bs to be multi linear.
R7BN<cr> Recorder output of 7Bs to be linear.
R7BT<cr> Recorder output of 7Bs to be B7B2 traditional.
R7Ex<cr> Recorder output of 7Es to be linear to 10-x Torr
where 7 => x => 2 and 10V is at 10-x Torr
R7FM<cr> Recorder output of 7Fs to be multi linear.
1V/Decade and 9V is at 10-2 Torr.
R7FL<cr> Recorder output of 7Fs to be logarithmic. 9V is at 10-2 Torr.
R7Fx <cr> Recorder output of 7Fs to be linear to 10-x Torr
Where 8 => x >= 2 and 10V is at 10-x Torr.
R7EL<cr> Recorder output of 7Es to be logarithmic.
R7EM<cr> Recorder output of 7Es to be multi-linear.
SET POINTS
CPn<cr> Clear setpoint values for relay "n".
PCA<cr> Control all relays via RS232.
PCn<cr> Control relay "n" via RS232.
PFn<cr> Turn off controlled relay "n" via RS232.
PNn<cr> Turn on controlled relay "n" via RS232.
PUA<cr> Remove all relays from RS232 control.
PUn Remove relay "n" from RS232 control.
SAxSy Assign relay x to station y.
SPx<cr> Send the station assignment of relay x.
SPxN<cr> Send the "ON" setting of relay x.
SPxF<cr> Send the "OFF" setting of relay x.
SSxFyyyyu<cr> Set relay #x "OFF" to yyyy, units H or L.
SSxNyyyyu<cr> Set relay #x "ON" to yyyy, units H or L.

Table of Contents