EasyManuals Logo

TESTO 184 User Manual

TESTO 184
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
4 Product description
16
Display T2 T3 H1 G1
Maximum acceleration reading , X axis (Max,
x, g)
- - - X
Maximum acceleration reading, Y axis (Max,
y, g)
- - - X
Maximum acceleration reading, Z axis (Max,
z, g)
- - - X
Time mark ( )
X X X X
Battery life in days ( )
X X X X
STOP key
Rec mode and stop criterion Button Stop programmed.
> Press STOP key for approximately 3 seconds to end the
measurement program.
- The measurement program ends: Mode status LED flashes red,
End appears on the display.
START + STOP key
The data loggers may be placed in hibernation mode to prolong the
battery life. In this mode, the status LEDs and the display are
disabled.
WAIT or End operating mode.
> Press the START key and the STOP key simultaneously for
approximately 3 seconds.
- Hibernation mode is activated.
4.4. Important information and glossary of terms
Single-use data logger (versions T1 and T2): the data logger
has a time-limited service life, which begins from the first time
the program is started.
Start and stop setting: the criteria for starting and stopping the
program are defined in the configuration file.
One of the criteria must be selected for starting the program.
When selecting the criterion, a time delay can be entered
(program starts x minutes after pressing the key).
Both criteria can also be selected for stopping the program. The
criterion that occurs first stops the program.
Measuring interval: the measurement interval defines the
intervals at which readings are saved.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO 184 and is the answer not in the manual?

TESTO 184 Specifications

General IconGeneral
Temperature accuracy±0.5 °C
Resolution temperature0.1 °C
Humidity measuring range0 to 100 %RH
Humidity accuracy±3 %RH
Resolution humidity0.1 %RH
Measuring rate1 min to 24 h
Protection classIP67
Battery lifeApprox. 2 years

Related product manuals