5 Using the product
19
5 Using the product
5.1. Configuring the data logger
Showing/changing the configuration
Adobe Reader software (version X or later) is required.
The data logger must not be in Rec mode.
1. Connect the data logger to a PC via the USB port.
- The status LEDs are disabled, uSb is displayed (instruments
with display). The device drivers are installed automatically.
- The window Automatic playback is displayed.
2. Click on Open folder to view files.
- The file explorer opens.
3. Open the file testo 184 configuration.pdf.
4. Make changes to the configuration. Please note:
The type of instrument used must be set correctly.
Existing configuration data can be imported by clicking on
the Import button. The configuration data to be imported
must be available in XML data format.
When using the Configuration Assistant, some functions are
predefined or filled in automatically. Expert Mode must be
enabled in order to use and manually set all instrument
functions.
5. Export changes to the configuration by clicking on the
button on the data logger.
- A window opens for exporting form data.
6. Select the data logger as the storage location
(Drive TESTO 184) and export the configuration data by
clicking on the Save button.
- The configuration is stored on the data logger as an XML file.
The XML file can be used as a template for other data loggers (via
the import function in the configuration PDF)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO 184 and is the answer not in the manual?

TESTO 184 Specifications

General IconGeneral
Temperature accuracy±0.5 °C
Resolution temperature0.1 °C
Humidity measuring range0 to 100 %RH
Humidity accuracy±3 %RH
Resolution humidity0.1 %RH
Measuring rate1 min to 24 h
Protection classIP67
Battery lifeApprox. 2 years