7 Tips and assistance
25
7 Tips and assistance
Questions and answers
Question
Possible causes/solution
E0x is displayed
(instruments with display),
all status LEDs are flashing
red
An error has occurred.
E01: Configuration unsuccessful /
PDF file faulty.
E02, E03, E04 or E05: sensor
faulty.
E06: maximum number of time
marks set, new time mark cannot
be set.
---- is displayed
(instruments with display)
No reading available (after setting
a time mark)
Reading invalid.
Err is displayed
(instruments with display)
Configuration not possible,
e.g. because
Rec
mode is active.
Configuration via the PDF
file is not possible
If you have used the Comfort Software
21 CFR Part 11 for the configuration,
configuration via the PDF file is
disabled.
The size of the PDF
configuration file is 0 kB or
the file is damaged.
Copy the PDF file of another testo
184
USB transport data logger or
download the configuration file from
the Testo website:
http://www.testo.com/.
Different time or time zone
in the report
> The testo 184 USB transport data
logger has not been configured
after a battery replacement.
Repeat the configuration to restore
the correct time settings.
> Check whether the PC which was
used for the configuration has the
correct time settings.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO 184 and is the answer not in the manual?

TESTO 184 Specifications

General IconGeneral
Temperature accuracy±0.5 °C
Resolution temperature0.1 °C
Humidity measuring range0 to 100 %RH
Humidity accuracy±3 %RH
Resolution humidity0.1 %RH
Measuring rate1 min to 24 h
Protection classIP67
Battery lifeApprox. 2 years