3 Specifications
5
3 Specifications
3.1. Use
The testo 184 USB transport data loggers are used to save and
read out individual readings and measurement series. They have
been specifically designed for monitoring the transportation of
products subject to cold chain requirements.
Temperature and humidity readings are saved throughout the
entire duration of the measurement program.
Acceleration readings are monitored throughout the duration of the
measurement program and saved when the set limit value is
exceeded.
Data logger programming and measurement report output are
implemented via PDF files, no software needs to be installed.
Product versions T1 and T2 are single-use data loggers with a
time-limited service life.
Configuration and readout of the testo 184 G1 USB
transport data logger are not possible with the testo
Saveris CFR Software and its transport add-ons.
3.2. Technical data
testo 184 T1
Feature Values
Probe type internal NTC temperature sensor
Measuremen
t channels
1 internal
Measuremen
t parameters
[unit]
Temperature [°C, °F]
Measuring
range
-35 to 70 °C
Accuracy ±0.5 K
Resolution 0.1 °C
Operating
temperature
-35 to 70 °C